Data

SCPI Commands

SENSe:CORRection:LOSS:CALibration:TABLe:DATA
class DataCls[source]

Data commands group definition. 1 total commands, 0 Subgroups, 1 group commands

class DataStruct[source]

Response structure. Fields:

  • Frequency: List[float]: Frequency of the measurement point. Range: 0 Hz to 999.99 GHz, Unit: HZ

  • Loss: List[float]: Loss of the measurement point. Range: -999.99 GHz to 999.99 GHz, Unit: DB

get() DataStruct[source]
# SCPI: [SENSe]:CORRection:LOSS:CALibration:TABLe[:DATA]
value: DataStruct = driver.applications.k30NoiseFigure.sense.correction.loss.calibration.table.data.get()

This command defines the contents of the currently selected calibration loss table. Each entry of the loss table consists of one measurement point and the corresponding loss. The table can contain up to 10001 entries. If you create a new table with this command, it overwrites the current entries of the loss table.

return

structure: for return value, see the help for DataStruct structure arguments.

set(frequency: List[float], loss: List[float]) None[source]
# SCPI: [SENSe]:CORRection:LOSS:CALibration:TABLe[:DATA]
driver.applications.k30NoiseFigure.sense.correction.loss.calibration.table.data.set(frequency = [1.1, 2.2, 3.3], loss = [1.1, 2.2, 3.3])

This command defines the contents of the currently selected calibration loss table. Each entry of the loss table consists of one measurement point and the corresponding loss. The table can contain up to 10001 entries. If you create a new table with this command, it overwrites the current entries of the loss table.

param frequency

Frequency of the measurement point. Range: 0 Hz to 999.99 GHz, Unit: HZ

param loss

Loss of the measurement point. Range: -999.99 GHz to 999.99 GHz, Unit: DB