Otype
SCPI Commands
OUTPut:TRIGger<TriggerPort>:OTYPe
- class OtypeCls[source]
Otype commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(triggerPort=TriggerPort.Default) TriggerOutType [source]
# SCPI: OUTPut:TRIGger<tp>:OTYPe value: enums.TriggerOutType = driver.applications.k60Transient.output.trigger.otype.get(triggerPort = repcap.TriggerPort.Default)
This command selects the type of signal generated at the trigger output.
- param triggerPort
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)
- return
output_type: DEVice Sends a trigger signal when the R&S FSWP has triggered internally. TARMed Sends a trigger signal when the trigger is armed and ready for an external trigger event. UDEFined Sends a user-defined trigger signal. For more information, see method RsFswp.Applications.K30_NoiseFigure.Output.Trigger.Level.set.
- set(output_type: TriggerOutType, triggerPort=TriggerPort.Default) None [source]
# SCPI: OUTPut:TRIGger<tp>:OTYPe driver.applications.k60Transient.output.trigger.otype.set(output_type = enums.TriggerOutType.DEVice, triggerPort = repcap.TriggerPort.Default)
This command selects the type of signal generated at the trigger output.
- param output_type
DEVice Sends a trigger signal when the R&S FSWP has triggered internally. TARMed Sends a trigger signal when the trigger is armed and ready for an external trigger event. UDEFined Sends a user-defined trigger signal. For more information, see method RsFswp.Applications.K30_NoiseFigure.Output.Trigger.Level.set.
- param triggerPort
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)