Data

SCPI Commands

CALCulate<Window>:LIMit<LimitIx>:UPPer:DATA
class DataCls[source]

Data commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get(window=Window.Default, limitIx=LimitIx.Default) List[float][source]
# SCPI: CALCulate<n>:LIMit<li>:UPPer[:DATA]
value: List[float] = driver.calculate.limit.upper.data.get(window = repcap.Window.Default, limitIx = repcap.LimitIx.Default)

This command defines the vertical definition points of an upper limit line.

param window

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)

param limitIx

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Limit’)

return

limit_line_points: Variable number of level values. Note that the number of vertical values has to be the same as the number of horizontal values set with method RsFswp.Calculate.Limit.Control.Data.set. If not, the R&S FSWP either adds missing values or ignores surplus values. Unit: DBM

set(limit_line_points: List[float], window=Window.Default, limitIx=LimitIx.Default) None[source]
# SCPI: CALCulate<n>:LIMit<li>:UPPer[:DATA]
driver.calculate.limit.upper.data.set(limit_line_points = [1.1, 2.2, 3.3], window = repcap.Window.Default, limitIx = repcap.LimitIx.Default)

This command defines the vertical definition points of an upper limit line.

param limit_line_points

Variable number of level values. Note that the number of vertical values has to be the same as the number of horizontal values set with method RsFswp.Calculate.Limit.Control.Data.set. If not, the R&S FSWP either adds missing values or ignores surplus values. Unit: DBM

param window

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)

param limitIx

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Limit’)