Select

SCPI Commands

DIAGnostic:SERVice:INPut:SELect
class SelectCls[source]

Select commands group definition. 1 total commands, 0 Subgroups, 1 group commands

set(signal: DiagnosticSignal) None[source]
# SCPI: DIAGnostic:SERVice:INPut[:SELect]
driver.diagnostic.service.inputPy.select.set(signal = enums.DiagnosticSignal.AIQ)

This command activates or deactivates the use of an internal calibration signal as input for the R&S FSWP.

param signal

CALibration Uses the calibration signal as RF input. MCALibration Uses the calibration signal for the microwave range as RF input. RF Uses the signal from the RF input. SYNTtwo Uses the calibration signal to check the phase noise of the two synthesizers. A second synthesizer is available as an hardware option.