Result
SCPI Commands
SENSe:ADEMod:FM:AFSPectrum:RESult
- class ResultCls[source]
Result commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(trace_mode: TraceModeB) float [source]
# SCPI: [SENSe]:ADEMod:FM:AFSPectrum:RESult value: float = driver.sense.ademod.fm.afSpectrum.result.get(trace_mode = enums.TraceModeB.AVERage)
This command reads the result data of the evaluated signal in the specified trace mode. The data format of the output data block is defined by the FORMat command (see method RsFswp.FormatPy.Data.set) . The trace results are configured for a specific evaluation. The following table indicates which command syntax refers to which evaluation method, as well as the output unit of the results.
Table Header: Command syntax / Evaluation method / Output unit
ACV[:TDOMain] / AC-Video time domain / V
ACV:AFSpectrum / AC-Video spectrum / V
AM[:ABSolute][:TDOMain] / RF time domain / dBm
AM:RELative[:TDOMain] / AM time domain / %
AM:RELative:AFSPectrum / AM spectrum / %
FM[:TDOMain] / FM time domain / kHz
FM:AFSPectrum / FM spectrum / kHz
PM[:TDOMain] / PM time domain / rad or °
PM:AFSPectrum / PM spectrum / rad or °
SPECtrum / RF spectrum / dBm (logarithmic display) or V (linear display) .
- param trace_mode
WRITe | AVERage | MAXHold | MINHold
- return
trace_mode_result: The specified trace mode must be one of those configured by SENS:ADEM:Evaluation:TYPE, see [SENSe:]ADEMod:SPECtrum[:TYPE]. Otherwise a query error is generated.