Upper

SCPI Commands

SENSe:ADJust:CONFigure:HYSTeresis:UPPer
class UpperCls[source]

Upper commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get() float[source]
# SCPI: [SENSe]:ADJust:CONFigure:HYSTeresis:UPPer
value: float = driver.sense.adjust.configure.hysteresis.upper.get()

When the reference level is adjusted automatically using the [SENSe:]ADJust:LEVel command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines an upper threshold the signal must exceed (compared to the last measurement) before the reference level is adapted automatically.

return

threshold: Range: 0 dB to 200 dB, Unit: dB

set(threshold: float) None[source]
# SCPI: [SENSe]:ADJust:CONFigure:HYSTeresis:UPPer
driver.sense.adjust.configure.hysteresis.upper.set(threshold = 1.0)

When the reference level is adjusted automatically using the [SENSe:]ADJust:LEVel command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines an upper threshold the signal must exceed (compared to the last measurement) before the reference level is adapted automatically.

param threshold

Range: 0 dB to 200 dB, Unit: dB