Data

SCPI Commands

FORMAT REAL,32;TRACe:IQ:DATA
class DataCls[source]

Data commands group definition. 3 total commands, 2 Subgroups, 1 group commands

get() List[float][source]
# SCPI: TRACe:IQ:DATA
value: List[float] = driver.trace.iq.data.get()

This command initiates a measurement with the current settings and returns the captured data from I/Q measurements. This command corresponds to: INIT:IMM;*WAI;method RsFswp.Trace.Iq.Data.Memory.get_ However, the method RsFswp.Trace.Iq.Data. get_ command is quicker in comparison. Note: Using the command with the *RST values for the TRACe:IQ:SET command, the following minimum buffer sizes for the response data are recommended: ASCII format 10 kBytes, binary format: 2 kBytes

return

iq_data: Measured voltage for I and Q component for each sample that has been captured during the measurement. The number of samples depends on TRACe:IQ:SET. In ASCII format, the number of results is 2* the number of samples. The data format depends on method RsFswp.Applications.IqAnalyzer.Trace.Iq.Data.FormatPy.set. Unit: V

Cloning the Group

# Create a clone of the original group, that exists independently
group2 = driver.trace.iq.data.clone()

Subgroups