SymbolRate
SCPI Commands
TRACe:IQ:SRATe
- class SymbolRateCls[source]
SymbolRate commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get() int [source]
# SCPI: TRACe:IQ:SRATe value: int = driver.applications.iqAnalyzer.trace.iq.symbolRate.get()
This command sets the final user sample rate for the acquired I/Q data. Thus, the user sample rate can be modified without affecting the actual data capturing settings on the R&S FSWP. Note: The smaller the user sample rate, the smaller the usable I/Q bandwidth, see ‘Sample rate and maximum usable I/Q bandwidth for RF input’.
- return
count: No help available
- set(count: int) None [source]
# SCPI: TRACe:IQ:SRATe driver.applications.iqAnalyzer.trace.iq.symbolRate.set(count = 1)
This command sets the final user sample rate for the acquired I/Q data. Thus, the user sample rate can be modified without affecting the actual data capturing settings on the R&S FSWP. Note: The smaller the user sample rate, the smaller the usable I/Q bandwidth, see ‘Sample rate and maximum usable I/Q bandwidth for RF input’.
- param count
The valid sample rates are described in ‘Sample rate and maximum usable I/Q bandwidth for RF input’. Unit: HZ