Lower
SCPI Commands
SENSe:ADJust:CONFigure:HYSTeresis:LOWer
- class LowerCls[source]
Lower commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get() float [source]
# SCPI: [SENSe]:ADJust:CONFigure[:HYSTeresis]:LOWer value: float = driver.applications.k70Vsa.sense.adjust.configure.hysteresis.lower.get()
When the reference level is adjusted automatically using the [SENSe:]ADJust:LEVel command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines a lower threshold the signal must fall below (compared to the last measurement) before the reference level is adapted automatically.
- return
hysteresis_lower: No help available
- set(hysteresis_lower: float) None [source]
# SCPI: [SENSe]:ADJust:CONFigure[:HYSTeresis]:LOWer driver.applications.k70Vsa.sense.adjust.configure.hysteresis.lower.set(hysteresis_lower = 1.0)
When the reference level is adjusted automatically using the [SENSe:]ADJust:LEVel command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines a lower threshold the signal must fall below (compared to the last measurement) before the reference level is adapted automatically.
- param hysteresis_lower
Range: 0 dB to 200 dB, Unit: dB